Start of a program, E 36 – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 36

Advertising
background image

Menu functions

36

9.1. Start of a program

The start menu of the integrated programmer allows one of six previously
stored temperature programs to be called up and started. The profiles are
started manually or via the integrated timer.

Requirements:

1. Create a program. (refer to next chapter)

2. Return to the Start-MENU and confirm the desired setting of each

MENU item with the key

3. Set a start time (>TIME< >DATE< >YEAR<) if the program is to

be started by the internal timer.

>Program Number< Select the program that should be started.

0 ... 5

> at Step<

Select the Step at which the program should be
started. 0 … 59

> Runs <

Select the number of program repetitions. 1 ... 99

> Program End<

Status at program end .
StdBy, PSetp, Setp 1, Setp 2, Setp 3

> RUN <

Start time (now / time)

Press the

key if a parameter is to be retained. (prior to OK)

Level 2

Parameter /Value

• Set a parameter using the numeric

keypad and confirm by pressing

.

Possible parameters: 0 …5

• Set a parameter using the numeric

keypad and confirm by pressing

.

Possible parameters: 0 … 59

• Set a parameter using the numeric

keypad and confirm by pressing

.

Possible parameters: 1 … 99

Set desired parameters with

and

.

Possible parameters:
StdBy , PSetp, Setp 1, Setp 2 , Setp 3

Advertising