Status at the end of a program, Indication after successful start – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 38

Advertising
background image

Menu functions

38

9.1.1.

Status at the end of a program


Program end

StdBy , PSetp, Setp 1, Setp 2 , Setp 3

This parameter is used to determine if the unit switches to the -OFF-
condition at the end of a profile or if temperature control is continued,
and the working temperature setpoint that is to be used if
temperature control is continued.

StdBy – If the end of the profile is reached, the circulator switches to
the -OFF- condition. The VFD display indicates the working
temperature setpoint that was displayed in line 2 before the program
was started (example: S 25.00 °C ).

PSetp –(Setpoint of the ProGrammer) At the end of the program, the
setpoint of the last program step is written to the position of setpoint 3
(example: 80.00 °C) and indicated in line 1 of the LCD display. The
circulator operates and controls to this temperature until the key

is pressed or a new profile is started.


Setp 1, Setp 2, Setp 3 – (“Setp” short for Setpoint)
Before starting the programmer, a temperature value is set for one of
the three setpoints in the

menu.

(example: setpoint 2 58.00 °C)
The circulator operates and controls to this temperature at the end of
the profile until the key

is pressed or a new program is started.

9.1.2.

Indication after successful start

Examples:

The started programmer displays the currently calculated setpoint in line
1. The value changes until the final temperature of the section is reached
(temperature ramp).

1

st

line:

Setpoint of the programmer

2

nd

line:

Actual temperature value

for internal control = IntAct: xxx.xx

for external control = ExtAct: xxx.xx

3

rd

line:

Selected profile and the current section (step).

4

th

line:

With the

keys, the following information can be

displayed in line 4.

• Remaining time of the current section (standard display)

• Total remaining time: profile x number of repetitions

(All: Days/hours:minutes:seconds)

Particularities :

1. If a total time of more than 999 days is calculated, only the hours

are displayed by the unit.
(All: hours:minutes:seconds)

2. If the time period of a section is edited after the start, then the

remaining total time can no longer be displayed.

Advertising