Example: 3-point calibration for internal control – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 64

Advertising
background image

Menu functions

64

9.7.4.

Example: 3-point calibration for internal control

In the temperature range from 80 °C to 160 °C the calibration curve of the
temperature sensor (TT) is to be adjusted to the actual temperatures at
measuring point (CT).

1. Set circulator to internal control:

The type of control can be set only in the –OFF- mode.

80.00 °C

120.00 °C

160.00 °C

2. Set working temperature setpoint

See standard display line 1

Example 80.00 °C

Press the keys

and

and confirm by pressing

.

The bath is heated up.
Wait for approx. 5 minutes until the temperature is constant.

3. Reading of temperature measuring device

Read the value of measuring point CT on the device and enter under menu
item > CalValue X< by using the keypad.

CalValue 1< (79.70 °C)
CalValue 2< (119.50 °C)
CalValue 3< (159.30 °C)

4. Calibration

Press the

key if a parameter is to be retained. (prior to OK)

Setting is required only for the first calibration point.

Level 2

Parameter-Level

Set sensor: >intern<.

• The parameter flashes, switch by

pressing

and

Set to >no<

• The parameter flashes, switch by

pressing

and

A >3-point< calibration is carried out.

• The parameter flashes, switch by

pressing

and

Advertising