Program administration, creation, E 41 – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 41

Advertising
background image

HL, SL

41

9.2. Program administration, creation

The integrated programmer permits fast and easy programming of
setpoint temperature sequences. This temperature sequence is called
program. A program is composed of individual sections (Step).
The sections are defined by duration (t) or Gradient (°/t) and target
temperature. The target temperature is the setpoint (SP), which is
achieved at the end of a section. The programmer calculates the
temperature ramp from the difference in time and temperature.

Create programs

A flashing segment indicates that a number needs to be entered.
Under the submenu “Edit”, enter a program number.
Six programs may be stored (numbers 0 to 5).

When the program is running, only sections with complete information for
the target temperature and time period are considered.
It makes sense to skip section numbers in the profile in order to use them
later for corrections to the profile.

Important:
If a time of 00:00 is set for a profile, the profile is continued with the next
section only after the setpoint temperature (±0.2 °C) is reached.

Example:
Step

(Nr.)

1

3

5

6

9

11

14

Setpoint

(°C)

50

50

100

100

80

80

20

Time

(h:m) 00:08 00.02 00:10 01:05 00:10 00:05 00:30

Advertising