Electrical connections

82

11. Electrical connections

Notice:

Use shielded cables only.
The shield of the connecting cable is electrically connected to the plug housing.

The unit ensures safe operation if connecting cables with a maximum length of 3 m are
used. The use of longer cables does not affect proper performance of the unit, however
external interferences may have a negative impact on safe operation (e.g. cellular
phones).

4

1

3

2

ext. Pt100

Socket for external Pt100 sensor

1

4

2

3

Shield

Plug

Look on

soldering side.

Pt100

Pin assignment:
Pin

Signal

1

I+

2

U+

3

U-

4

I-

The shield of the connecting cable is electrically connected to the plug housing
and the sensor tube.


9

6

5

1

SERIAL

RS232/RS485 serial interface
This port can be used to connect a computer with an RS232 or RS485 cable
for remote control of the circulator.

Pin assignments RS232:

Pin 2

RxD

Receive Data

Pin 3

TxD

Transmit Data

Pin 5

0 V

Signal GND

Pin 7

RTS

Request to send

Pin 8

CTS

Clear to send

Pin 1; 4; 6, 9 Reserved - do not use!

Pin assignments RS485:
Pin 3

B

Pin 5

0 V

Signal GND

Pin 6

+5 V (max. 50 mA)

Pin 8

A


Pin 1; 2; 4; 7; 9 Reserved - do not use!

Notice:

RS485 serial interface

The max. load current between pin 5 and pin 6 is 50 mA.