Alarm-output – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 78

Advertising
background image

Menu functions

78

9.9.5.

Alarm-output

ALARM


Alarm output (10)

(for external alarm signal)

This socket is a potential-free change-over contact. With the
adjustments in the menu >Inputs/Outputs<, all operating conditions
can be signaled without having to change the pin assignments.

Meaning of the terms under menu item >Function< :

The circulator is in condition

>StandBy<

or

>Alarm<

For >Type: normal<, pins 2 and 3 are connected in any case
according to the selected >Function<.

For >Type: inverse<, pins 2 and 1 are connected in any case
according to the selected >Function<.

Switching capacity

max. 30 W/40 VA

Switching voltage

max. 125 V

∼/−

Switching current

max. 1 A

Advertising