List of commands – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 86

Advertising
background image

Remote control

86

The out commands are valid only in remote control mode.

Examples:

When the RS485 interface is used, the three-digit instrument address
precedes each command.
(example: address Ad32 = A0 32 )

Command to set the working temperature >Setpoint1< to 55.5 °C

out_sp_00

55.5

A032_out_sp_00

55.5

Command to retrieve the working temperature >Setpoint1<

in_sp_00

A032_in_sp_00

Response from the circulator:

55.5

LF

A032_55.5

LF

Notice:

out commands

Temperature values can be displayed on the circulator in °C or °F.
The settings sent via the interface must also be made in the selected unit according
to this setting.


12.3. List of commands

out commands: Setting parameters or temperature values.

Command

Parameter

Response of circulator

out_mode_01

0

Use working temperature >Setpoint1<

out_mode_01

1

Use working temperature >Setpoint2<

out_mode_01

2

Use working temperature >Setpoint3<

out_mode_02

0

Selftuning “off”.
Temperature control using the stored parameters.

out_mode_02

1

Selftuning “once”
Single selftuning of the controlled system after the next start.

out_mode_02

2

Selftuning “always”
Continual selftuning of the controlled system whenever a new
setpoint is to be reached.

out_mode_03

0

Set external programmer input to voltage.

Voltage

0 V ... 10 V

out_mode_03

1

Set external programmer input to current.

Current

0 mA ... 20 mA

out_mode_04

0

Temperature control of internal bath.

out_mode_04

1

External control with Pt100 sensor.

out_mode_05

0

Stop the unit = R –OFF-.

Advertising