Important: Your station setup will determine whether you have to move in the positive or negative
direction.

Switch to Force Control

Switching to force control helps ensure that you do not impart excessive force on the specimen during
installation. Once in force control, the controller will follow your positioning commands while working to
maintain the force that you specify—in this case 0.00 kN.

1. In the Station Manager main window, click the Manual Command icon.

2. Set Control Mode to Force.

3. Set Manual Cmd to 0.00 kN.

Clamp the Specimen in the Upper Grip

Using the limits that you set earlier and having the load frame in force control during clamping help ensure
that excessive force is not imparted on the specimen during clamping. Nevertheless, there may be times
when you have to recover from a tripped limit that occurs during the clamping process. Instructions for
that situation are provided below.

1. Clamp the specimen in the upper grip.

84 | MTS Landmark Test System Operation