Do not locate above heat emitting objects.

Sensor Installation Environment

Install the sensor at a location that satisfies the following environmental and
installation conditions.

Environmental Conditions

Installation Conditions

Do not install at a location exposed to rain or other moisture. Moisture can

lead to reduced sensitivity of the sensor, and reduced insulation.

A target surface area of not less than three times the tip diameter centered

on the sensor is required, as illustrated below.

Feature Specification

Ambient temperature

ATEX applications must be in a range of -35…80 °C (-31…176 °F) when devices
are operating.
CSA applications must be in a range of -35…85 °C (-31…185 °F) when devices
are operating.
Other applications must be in a range of -35…177 °C (-31…350 °F) when devices
are operating

Ambient humidity

Must be in a range of 30…95% RH (noncondensing) when devices are operating.

Vibrational conditions

Must be 10 m/s

2

(1 g) or less at 10…150 Hz. (If the sensor vibrates, an accurate

measurement cannot be made.)

Air cleanliness

We recommend an air dust-particle amount of 0.2 mg/m

3

or less.

We recommend an especially low amount of corrosive gasses, such as hydrogen
sulfide, NOx gas, and chlorine, and conductive particles, such as iron dust and
carbon. The allowable amount sof hydrogen sulfide and NOx gas, based on JEIDA-
29 (1979) Class S1, are shown below.

JEIDA: Japanese Electronic Industry Development Association

JEIDA-29 (1979) CLASS S1
Hydrogen sulfide: 0.01 ppm or less, NOx gas: 0.05 ppm or less
(Ambient temperature: 25 °C ± 5 °C (77 °F ± 9 °F), humidity: 40…80%
RH)