Connect the module, Verify the connections, Set gap voltage – Rockwell Automation 1442 Eddy Current Probe System User Manual

Page 35

Advertising
background image

Rockwell Automation Publication ICM-UM004C-EN-E - February 2014

35

Installation

Chapter 1

Connect the Module

The 1442 sensors can be connected to many different Allen-Bradley 1440 XM
Series or 1444 Dynamix® Series modules. Refer to the appropriate Module User
Manual for wiring requirements and instructions on how to wire the sensor to the
module.

Verify the Connections

Before turning on the power, verify the following connections:

· Be sure that there are no loose terminals, and that all wiring is properly

connected.

· Check that the power line for the power source is connected to

NEGATIVE PWR (-24V) on the measurement module or its terminal
base.

· Be sure that the driver and sensor are installed at locations where the

installation environmental conditions are satisfied.

· Be sure that there are no problems with the driver and sensor installation,

and they are

not installed at the following types of locations:

Locations with high temperatures and high humidity.
Locations with dust.
Location exposed to vibration.
Locations where there are metal objects, other than the target, near the

sensor.

After checking all items, check the set gap voltage values.

Set Gap Voltage

Perform confirmation of set gap voltage to maintain the performance of this unit
when doing the following:

Supplying power to the unit for the first time
More than one year has passed from the last confirmation
The performance of this unit has been reduced due to a problem of some

sort

Follow these steps to check the set gap voltage.

1. Turn on the power.
2. Allow the unit to warm up for 5 minutes to stabilize the output.