A - wire the unit to a monitor system, Appendix a, Wire the unit to a monitor system – Rockwell Automation 1442 Eddy Current Probe System User Manual

Page 51: Appendix

Advertising
background image

Rockwell Automation Publication ICM-UM004C-EN-E - February 2014

51

Appendix

A

Wire the Unit to a Monitor System

The 1442 Series Probe System is designed to satisfy the API-670 standard. Any
monitor designed to connect API-670 probes can be used with these sensors.

Consider the following recommendations when wiring the probe driver to a
monitor:

· Use a good quality instrumentation cable with three-conductor stranded

wire and shield.

Wire must be rated with a maximum capacitance of 60 pF/ft

(197 pF/m) and inductance of 0.3 μH/ft (1 μH/m).

Use wire with insulation suitable for the environment and with

adequate tensile strength and flexibility for the application.

Use wire with a foil shield for use in environments where radio

frequency interference (RFI) may be present. Use a wire with a braid
shield for environments where electromagnetic interference (EMI) may
be present.

Use 0.75…1.25 mm2 (18…16 AWG) gauge wire.

· Make sure the wire is isolated from power cables and any other wiring that

may be transmitting high-voltage power or control signals.

· Any cable transmitting pulse-type vibration signals such as a phase marker

or speed pulse must be isolated from displacement and vibration signals.

· Run wire within conduit and cable trays and as per any local electrical

codes.

· Do not exceed a wire length of 500 m (546.81 yds). However, limiting the

length to 300 m (328.08 yds) transmits vibration signals in the 0…10 kHz
frequency range with minimal attenuation. When longer lengths are
needed the capacitance of the cable and the desired frequency response of
the system must be considered.

· In most cases, ground the cable shield at only one point, generally at the

monitor.

Advertising