2 Installation

845 Eluent Synthesizer / Instructions for Use 8.845. 8001EN

14

2 Connect capillary to the components

Components from a bottle:

• Insert 6.1803.130 PTFE tubing 24 through the bottle attach-

ment into the component bottle (e.g. the combination

6.1602.160 bottle attachment GL 45 and 6.1608.030 bottle;

see Optional accessories section 5.2.4).

Components from an MPak (from Chata Biosystems, see section

5.2.5):

• Push the 6.1808.200 MPak coupling onto 6.1803.130 PTFE

tubing 24 by loosening the screw cap from the coupling

and pushing the smaller opening over the tubing. Then

screw the coupling into the cap:

• Click the 6.1808.200 MPak coupling onto the lower outlet of

the MPak.

2.6 pH

Controller

The pH of the prepared mixture can, if necessary, be adjusted auto-

matically by using the (optional) extension kit. If the pH is to be adjusted

from one side then one pH Controller is required, adjustment from both

sides (without changing the titrant solution) requires two pH Controllers.

We recommend the use of

702 Titrinos as pH Controllers (for supply

package see Section 5.2.3). Other Titrinos can also be used (supply

package for required accessories see Section 5.2.3).

2.6.1

Tubing connection for pH Controller

If the pH is to be adjusted automatically after the mixing process then

adaptations are required to the tubing connections to the 845 Eluent

Synthesizer.

1 Replace bottle adapter 21

6.1602.180 Bottle adapter 21 must be replaced by 6.1602.190

Eluent bottle head.

• Screw off the 6.1838.000 FEP tubing connection(s) 20 from

6.1602.180 Bottle adapter 21.

• Screw the 6.1838.000 FEP tubing connection(s) 20 to M6

thread(s) of 6.1602.190 Eluent bottle head.