Metrohm 845 Eluent Synthesizer User Manual

Page 28

Advertising
background image

3 «Mix Control» software

845 Eluent Synthesizer / Instructions for Use 8.845. 8001EN

22

No [ 1 ; display only ]

Number of the Component (see section 3.4.2).

Component

Name of the Component.

pH-Adjustment

none

No pH adjustment will take place.

with request

pH adjustment will take place. It can be carried out
manually or automatically (702 Titrinos from Metrohm
are recommended). After pH adjustment the

LIVE DIS-

PLAY

window opens in which the volume added for pH

adjustment should be entered. This added volume is
taken into account
in the volume calculations of the
mixing process.

The titration volume should be below 10 mL. Titration
volumes above 10 mL result in an increased

Total

Volume.

autonomous

pH adjustment will take place. It will be carried out
automatically (702 Titrinos from Metrohm are recom-
mended). A pH-Controller must be selected. The
added volume will not be taken into account in the
volume calculations of the mixing process.

pH-Controller

(see section 2.6)

none

pH adjustment is carried out manually (must not be se-
lected if

autonomous

has been selected for

pH-Adjust-

ment

).

pH-Controller A

The pH is adjusted with the instrument that is connected
with the cable end marked A (of the two-headed
6.2141.210 Cable).

pH-Controller B

The pH is adjusted with the instrument that is connected
with the cable end marked B (of the two-headed
6.2141.210 Cable).

Advertising