4.3 Troubleshooting

845 Eluent Synthesizer / Instructions for Use 8.845. 8001EN

49

4.3 Troubleshooting

Problem

Cause

Remedy

Liquids escaping from
drainage tubing or
from bottom of rear
panel.

• Leaking connections

• Leak in instrument

1. Stop the current process

2. Remove rear panel

19 and

check the connections and dos-
ing units (see Section 4.2)

Components are not
aspirated.

• Leaking or blocked con-

nections

• Dosing units/Dosinos

defect

1. Stop the current process

2. Check the connections outside

the instrument

3. Remove rear panel

19 and

check the connections and Dos-
inos/Dosing units (see Section

4.2)

LED "Power" (1) does

not light up when in-

strument is connected.

• Mains cable defect

• Internal power supply

defect

• Replace the mains cable

• Contact Metrohm Service

Instrument not recog-
nized at software start-
up (LED "Power" (1)

lights up)

• Instrument not con-

nected to USB

• USB cable defect

• Connect instrument to USB of

the PC (see Section 2.3)

• Replace USB cable