2.6 pH Controller

845 Eluent Synthesizer / Instructions for Use 8.845. 8001EN

15

2 Install tubing connection to pH Controller

6.1838.010 FEP tubing connection should be used to make

the tubing connection to the pH Controller.
• Insert the black O-ring belonging to the 6.1838.010 FEP

tubing connection into an M6 thread of 6.1602.190 Eluent

bottle head.

• Then push the tip of the 6.1838.010 FEP tubing connection

through the O-Ring and M6 thread.

• Attach the 6.2726.090 Antidiffusion micro-valve to the tip of

the 6.1838.010 FEP tubing connection.

• Connect the other end of the 6.1838.010 FEP tubing con-

nection to the pH Controller (see Instructions for Use – 806

Exchange Unit).

3 Screw on bottle adapter

• Now screw 6.1602.190 Eluent bottle head onto eluent bottle

22.

4 Insert the Aquatrode in the Eluent bottle head

• Insert 6.0257.020 Aquatrode into the large opening of the

6.1602.190 Eluent bottle head. Watch out, that the Aquatro-

de looms deep enough into the mixture when the pH is

measured.

2.6.2

pH Controller cable connection

The methods at the pH Controller are started via a remote signal from

the 845 Eluent Synthesizer (see Section 3.3.3). Connect the pH Control-

ler to the 845 Eluent Synthesizer as follows:

1 Connect Remote Box

• Connect the 6.2148.010 Remote Box to connection "Re-

mote Box / MSB4" 15.

2 Connect pH Controller

• Connect 6.2141.210 Cable to the remote output of

6.2148.010 Remote Box.

• Connect the end of the 6.2141.210 Cable marked A to the

remote input of the pH Controller.

• Connect the end of the 6.2141.210 Cable marked B to the

remote input of the second pH Controller.

The Metrohm 702 Titrino can be used with the

SET method to set the

pH to a defined value (see the Instructions for Use of the correspon-

ding Titrino).

The method for the pH-adjustment and the corresponding parameters

have to be defined on the pH-Controller. The pH-adjustment is started

by a remote signal sent from the 845 Eluent Synthesizer to the pH-

Controller. For the installation and operation of the pH Controllers

please refer to the corresponding Instructions for Use.