5 ambient temperature, 6 dosing drive, 7 dosing unit – Metrohm 845 Eluent Synthesizer User Manual

Page 57: 8 reference conditions, 9 dimensions, Ambient temperature, Dosing drive, Dosing unit, Reference conditions, Dimensions

Advertising
background image

5.1 Technical data

845 Eluent Synthesizer / Instructions for Use 8.845. 8001EN

51

- EN/IEC 61000-4-2

- EN/IEC 61000-4-3

- EN/IEC 61000-4-4

- EN/IEC 61000-4-5

- EN/IEC 61000-4-6

- EN/IEC 61000-4-11

- EN/IEC 61000-4-14

5.1.5 Ambient

temperature

Nominal working range

Automatic inner temperature

monitoring

Storage

Transport

+5 C...+45 C (at max. 85% relative humidity)

> 70°C initial alarm

> 75°C alarm

–20°C...+60 °C

–40°C...+60 °C

5.1.6 Dosing

drive

Resolution

Dosing/Filling rates

10'000 increments per cylinder volume

18 seconds each per cylinder volume

5.1.7 Dosing

Unit

Volumes of cylinders
Chemical resistance of the

PETG housing

50 mL
Good resistance to:

aqueous solutions, dilute acids, alcohols and hydro-

carbons
Limited resistance to:

conc. organic acids, dilute aqueous alkalis (stress

cracks), acetone, isopropanol, tetrahydrofuran, hot

water
Not resistant to:

concentrated inorganic acids and bases, bromine,

chlorinated solvents, phenol, water vapor >100 °C

5.1.8 Reference

conditions

Ambient temperature

Relative humidity

Instrument status

+25 °C (±3 °C)

≤ 60%

In operation for at least 30 min

5.1.9 Dimensions

Housing material

Polyurethane

Width
Height
Depth

314 mm
378 mm
280mm

Weight

2.844.0010
2.844.0020


6920 g
7639 g

Advertising