2 fuses, 3 mains cable and mains connection, Mains cable – Metrohm 813 Compact Autosampler User Manual

Page 14: Mains connection, 3 mains cable and mains connection mains cable

Advertising
background image

2.2 Mains connection

8

813 Compact Autosampler


23

22
21

22

0

- 24
0 V

10

0

- 12

0 V

22

0

- 24

0 V

10

0

- 12
0 V

Fig. 4 Setting the mains voltage

2.2.2 Fuses

One of the two fuses 0.5 A/slow-blow for 100…120 V or 0.25

A/slow-blow for 220…240 V is installed in fuse holder

21

of the

813 Compact Autosampler as standard

Ensure that the instrument is never put into operation with fuses

of another type, otherwise there is danger of fire!

For checking or changing fuses, proceed as described in section

2.2.2.

2.2.3 Mains cable and mains connection

Mains cable

The instrument is supplied with one of three mains cables

• 6.2122.020 with plug SEV 12 (Switzerland, …)
• 6.2122.040 with plug CEE(7), VII (Germany, …)
• 6.2133.070 with plug NEMA 5-15 (USA, …)

which are three-cored and fitted with a plug with an earthing pin.

If a different plug has to be fitted, the yellow/green lead (IEC

standard) must be connected to protective earth (protection class

1).

Any break in the earthing inside or outside the instrument can

make it a hazard!

Mains connection

Plug the mains cable into mains connection plug

22

of the 813

Compact Autosampler (see Fig. 4).

220 – 240 V

100 – 120 V

21 Fuse holder

22

Mains connec-

tion plug

23

Mains switch