5 the metrohm ic control software, 1 operational conditions – Metrohm 813 Compact Autosampler User Manual

Page 33

Advertising
background image

3

Operation

813 Compact Autosampler

27

3.5 The Metrohm IC control software

To use the 813 Compact Autosampler with a 761 Compact IC or

a 7

90 Personal IC one has to create a time program. First of all the

IC has to send a Remote signal (Remote line 3) to synchronize

the command sequence (first Remote signal) of the 813 Compact

Autosampler. After that the injection valve is switched to 'Fill' posi-

tion. Th

en a second Remote signal starts the filling of the sample

loop. The pump duration for filling the sample loop is defined by

the third Remote signal. After sending this Remote signal the in-

jection valve is switched to 'Inject' position immediately which

starts the data aquisition. During sample injection the peristaltic

pump of the 813 Compact Autosampler is still running. 12 s after

receiving the Remote signal the Autosampler will stop the pump.

Fig. 17 Time program of the IC control software

3.5.1 Operational conditions

To ensure a correct interconnection (Autosampler – Ion chro-

matograph), the following conditions must be met:

• When the Ion chromatograph is operational the 813 Com-

pact Autosampler hast to be started before the 'Sample

Queue' of the IC.

• When using the 761 Compact IC the initial state of the Re-

mote line 3 has always to be set to 0 (set

System startup val-

ues

:

Remote line 3 = 0

).