4.1 Error Messages

30

813 Compact Autosampler

4 Appendix

4.1 Error Messages

An error occurring during a sample series is displayed by a blink-

ing red LED. The number of flashes indicates the error number. If

an error occurs during processing of a sample series, the

changer will then be switched into the 'HOLD' state. The current

method has to be halted by pressing <STOP>.

The list of possible error messages and their causes:

1* RAM defect

Call Metrohm-Service.

3* battery low

The battery for the permanent storage of the settings

must be replaced.

6* changer low power

The power supply cannot deliver enough power for the

simultaneous operation of all components currently in

use (pump and lift).

7* rack data missing

No sample rack is in position or no rack data can be

found for the sample rack that is in place.

8* invalid rack code

The rack code read by the changer could not be found

in the internal position tables.

10* raise lift first

Turning of a rack could not be carried out because the

lift was below the rest (shift) position.

12* changer overload

Load or resistance too large to turn the rack.