2.3 Attaching the accessories

12

813 Compact Autosampler

2.3.5 Placing the sample rack

1

2

Place sample rack

Place sample rack

24

on the

turntable of the 813 Compact

Autosampler acc. to Fig. 8.

Initialize rack

Press <RESET> to move the

rack to the home position, in

which the magnetic code can

be read (coding magnet 30).

30

Fig. 8 Sample rack placing

2.3.6 Adjusting the sample rack

If a new sample rack is placed on the 813 Compact Autosampler

for the first time, it must be adjusted. Make sure you have

mounted a steel needle or a PEEK needle before. Proceed as fol-

lows:

1 Place sample rack

Place empty sample rack

24

on the turntable of the 813 Com-

pact Autosampler and switch off the instrument.

2 Switch on the instrument

Keep the <HOLD> key pressed and simultaneously press the

mains switch on the right hand side of the

813 Compact Auto-

sampler. The instrument is initialized and the needle is placed

above rack position 1. Press the <START> key. The green LED

is now blinking twice.