3.6 Modular Metrohm IC System 732/733

The 813 Compact Autosampler can be controlled directly by a

Metrohm 732 IC Detector or by a PC control software, such as the

Metrohm IC Net 2.0 using the remote control lines of the 732 IC

Detector. See the 732 IC Detector user manual for details or read

the IC Net 2.0 documentation.

3.7 Metrohm 746 VA Trace Analyzer

Using a 813 Compact Autosamplers with a 746 VA Trace Ana-

lyzer require modified command sequences and monitoring set-

tings.

Between each sample determination a rinsing procedure for the

tubings has to be executed. For this purpose place a sample

tube containing rinsing solution on each second (even) rack posi-

tion. The samples to be determined have to be placed between

those rinsing tubes. This reduces the amount of a sample series

to 18 samples max.

3.7.1 Monitoring

The number of samples of a series is defined on the monitoring

page of the 746 VA Trace Analyzer, see below.

Auto.samples on 18 of 160
Auto.batch on 0 of 18
Start interval 0.0 h

3.7.2 Command sequence

A command sequence of the 746 VA Trace Analyzer has to con-

tain explicit trigger signals for the 813 Compact Autosampler that

are to be sent via the Remote output line 2. Use the 'SETCTRL'

command to send the required pulses, see line 4 and 5 of the ex-

ample method on the next page.