3 conditions to be met – Metrohm 813 Compact Autosampler User Manual

Page 35

Advertising
background image

3

Operation

813 Compact Autosampler

29

Sample method for the 746 VA Trace Analyzer:

Instructions t/s Main parameters Auxiliary parameters

------------- ----- ------------------------- ------------------------

1 STIR Rot.speed 2000 /min

2 PURGE

3 SMPL/M V.fraction mL V.total L

4 SETCTRL 1.0 Code *****1** Message

5 SETCTRL Code *****0** Message

6 REM 30.0 move to sample

7 SETCTRL 1.0 Code *****1** Message

8 SETCTRL Code *****0** Message

9 REM 300.0 transfer sample

10 SETCTRL 1.0 Code *****1** Message

11 SETCTRL Code *****0** Message

12 REM stop pump

13 DOS>M Soln.name Buffer V.add 2.000 mL

14 PURGE 90.0

15 (ADD

16 STIR Rot.speed 2000 /min

17 PURGE 30.0

18 (REP

19 SEGMENT Segm.name NiCoAdSV

20 REP)1

21 ADD>M Soln.name Standard V.add 0.010 mL

22 ADD)2

23 RINSE Cycles 1 V.rinse 20.000 mL

24 SETCTRL 1.0 Code *****1** Message

25 SETCTRL Code *****0** Message

26 REM 20.0 move to rinsing position

27 SETCTRL 1.0 Code *****1** Message

28 SETCTRL Code *****0** Message

29 REM 300.0 rinse tubes

30 SETCTRL 1.0 Code *****1** Message

31 SETCTRL Code *****0** Message

32 REM stop pump

33 RINSE Cycles 1 V.rinse 40.000 mL

34 RINSE Cycles 2 V.rinse 15.000 mL

35 END

3.7.3 Conditions to be met

For the correct intercommunication of the 813 Compact Auto-

sampler and the 746 VA Trace Analyzer the following conditions

must be met:

• Method 2 of the 813 Compact Autosampler has to be se-

lected.

• Always start the method of the 813 Compact Autosampler

first, then you can start the 746 VA Trace Analyzer.

• Ensure that the control line 2 (Remote output 2) of the 746

VA Trace Analyzer is reset (inactive state) before you start the

sample series.