1 typical combinations of instruments – Metrohm 760 Sample Changer User Manual
Page 15
2.4 Remote Connections
760 Sample Changer, Installation
12
14 lines (Output 0–13) are available for the emission of signals.
For receiving signals 8 lines are provided (Input 0–7). The
assignment of these lines is fixed for the 760 Sample Changer.
Pin assignment of the Remote socket:
Outpu t 6
Outpu t 7
Outpu t 4
Outpu t 2
Outpu t 0
Outpu t 8
Outpu t 9
Outpu t 10
Input 1
Input 3
Input 5
Input 7
Outpu t 11
0 Vo lt
+5 Volt
Output 5
Output 3
Output 1
Output 12
Output 13
Input 0
Input 2
Input 4
Input 6
0 Vo lt
The +5 V supply line may by charged with 20 mA maximally.
2.4.1 Typical Combinations of Instruments
760 Sample Changer — Titrino
with standard cable
B
A
Titrino
C
D
Remote control commands:
Output 0:
*************1
starts Titrino
Input 3:
****1***
awaits end of titration (EOD pulse)
Cable 6.2141.020
Inputs
Outputs
+5V
active = low
inactive = high
active = low