1 typical combinations of instruments – Metrohm 760 Sample Changer User Manual

Page 15

Advertising
background image

2.4 Remote Connections

760 Sample Changer, Installation

12

14 lines (Output 0–13) are available for the emission of signals.
For receiving signals 8 lines are provided (Input 0–7). The
assignment of these lines is fixed for the 760 Sample Changer.

Pin assignment of the Remote socket:

Outpu t 6
Outpu t 7
Outpu t 4
Outpu t 2
Outpu t 0
Outpu t 8
Outpu t 9
Outpu t 10
Input 1
Input 3
Input 5
Input 7
Outpu t 11

0 Vo lt

+5 Volt

Output 5
Output 3
Output 1

Output 12
Output 13

Input 0
Input 2
Input 4
Input 6

0 Vo lt



The +5 V supply line may by charged with 20 mA maximally.

2.4.1 Typical Combinations of Instruments

760 Sample Changer — Titrino

with standard cable

B

A

Titrino

C

D



Remote control commands:

Output 0:

*************1

starts Titrino

Input 3:

****1***

awaits end of titration (EOD pulse)

Cable 6.2141.020

Inputs

Outputs

+5V

active = low

inactive = high

active = low
inactive = high

Advertising