2.4 Remote Connections

760 Sample Changer, Installation

13

760 Sample Changer — Titrino — Dosimat 765/776





Remote control commands:

Output 0:

*************1

starts Titrino

Output 6:

*******1******

starts Dosimat

Input 3:

****1***

awaits end of titration (EOD pulse)



Pump Control via Remote Lines

Using methods 3 and 4, external pumps can be controlled via the
remote interface. To start a pump, an inverted pulse of duration of
at least 200 ms is sent to pin 7 (Output 9) of the remote socket.
To continue with the method the pump must indicate its readiness
by setting the input line 1 (pin 9) to low level for at least 20 ms.
Ask your pump distributor or Metrohm for customized connecting
cables.

Pin 9

Input

t

p

t

p

>20 ms



Pin 7

Output

t

p

t

p

>200 ms

V

CEO

= 40 V

I

C

= 20 mA



Cable 6.2141.040

+5V

active = low

inactive = high

active = low

inactive = high

B

A

Titrino

C

D

B

A

765

C

D