3.4 Configuration Procedure
760 Sample Changer, Introduction
18
The Configuration Procedure
Key
LEDs
Explanation
HOLD
+
"
Power ON"
green 1
∗
In order to start the configuration procedure the <HOLD> key
must be pressed while the sample changer is switched on.
The green LED lights up once to show that the configuration
procedure is active. The
method can now be selected.
HOLD
red 1-4
∗
green 1
∗
When the <HOLD> key is pressed the red LED blinks once
for the first method. Each time <HOLD> is pressed again the
next method is selected up to method 4, then method 1
again.
START
green 2
∗
<START> accepts the setting. The green LED blinks twice.
The
working position of the lift can now be set.
HOLD
red blinks
green 2
∗
When the <HOLD> key is pressed the lift can be moved
manually. Each further time the <HOLD> key is pressed the
direction in which the lift moves is altered. In this way the
working position can be set accurately.
START
green 3
∗
<START> accepts the setting. The working position is stored
specifically for the type of rack inserted. The green LED
blinks three times. The
rinsing position of the lift can now
be set.
HOLD
red blinks
green 3
∗
When the <HOLD> key is pressed the lift can be moved ma-
nually. Each further time the <HOLD> key is pressed the
direction in which the lift moves is altered.
START
green 4
∗
<START> accepts the settings. The rinsing position is stored
specifically for the type of rack inserted. The green LED
blinks four times. The
shift position of the lift can now be
set.
HOLD
red blinks
green 4
∗
When the <HOLD> key is pressed the lift can be moved ma-
nually. Each further time the <HOLD> key is pressed the
direction in which the lift moves is altered.
START
green 5
∗
<START> accepts the setting. The shift position is stored
specifically for the type of rack inserted. The green LED
blinks five times. The stirrer speed can now be set.