Metrohm 760 Sample Changer User Manual

Page 31

Advertising
background image

4.2 Method Listings

760 Sample Changer, Appendix

28

Methode 4

760 Sample Changer 760.0012
parameters
method 760_4
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 SAMPLE: - 1
3 SAMPLE: + 1
>sample sequence
1 MOVE 1 : sample
2 LIFT: 1 : work mm
3 STIR: * : ON s
4 CTL:Rm: START dos*
5 WAIT 10 s
6 CTL:Rm: START device1
7 SCN:Rm : End1
8 STIR: * : OFF s
9 LIFT: 1 : rinse mm
10 CTL:Rm: ****1*********
11 CTL:Rm: ****0*********
12 SCN:Rm : ******1*
13 LIFT: 1 : shift mm
14 WAIT 3 s
>final sequence
1 MOVE 1 : spec.1
2 LIFT: 1 : work mm
>changer settings
rack number 0
lift rate 1 25 mm/s
shift rate 20
shift direction: auto.
beaker test mode: single
on beaker error: MOVE
>stirring rates
stirrer 1 3
stirrer 2 3
>dosing unit def.
>manual stop
CTL Rmt: STOP device1
CTL RS232:
------------

¬ report header with program version

¬ method name
¬ number of samples (here full rack)

¬ initialize remote line
¬ initialize
sample position

¬ move next sample beaker to tower
¬ run lift with titration head to working position
¬ switch on stirrer
¬ start dosing device
¬ wait for 10 seconds
¬ start titrator
¬ wait for end of titration [EOD]
¬ switch off stirrer
¬ run lift with titration head to shift position
¬ switch on external pump
¬ reset remote line
¬ wait for ready signal of pump
¬ run lift to shift position
¬ wait for 3 seconds

¬ move special beaker to tower for conditioning
¬ run lift to work position to immerse sensor
------ settings of sampler functions --------------------




¬ on missing sample beaker, the next one is
selected automatically
------ stirring rates -----------------------------------------
¬ cannot be selected independently

------ settings for dosing devices ------------------------
------ action after manual stop -------------------------
¬ stop device 1



Advertising