4.2 Method Listings

760 Sample Changer, Appendix

26

Method 2

760 Sample Changer 760.0012
parameters
method 760_2
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 SAMPLE: - 1
3 SAMPLE: + 1
>sample sequence
1 SHIFTRATE: + 20
2 MOVE 1 : sample
3 LIFT: 1 : work mm
4 STIR: * : ON s
5 CTL:Rm: START dos*
6 WAIT 10 s
7 CTL:Rm: START device1
8 SCN:Rm : End1
9 STIR: * : OFF s
10 LIFT: 1 : shift mm
11 WAIT 3 s
12 SHIFTRATE: - 20
13 MOVE 1 : spec.1
14 LIFT: 1 : work mm
15 STIR: * : ON s
16 WAIT 5 s
17 STIR: * : OFF s
18 LIFT: 1 : shift mm
19 WAIT 3 s
>final sequence
1 MOVE 1 : spec.1
2 LIFT: 1 : work mm
>changer settings
rack number 0
lift rate 25 mm/s
shift rate 20
shift direction: auto.
beaker test mode: single
on beaker error: MOVE
>stirring rates
stirrer 1 3
stirrer 2 3
>dosing unit def.
>manual stop
CTL Rmt: STOP device1
CTL RS232:
------------

¬ report header with program version

¬ method name
¬ number of samples (here full rack)

¬ initialize remote line
¬ initialize
sample position

¬ shift direction to increasing numbers
¬ move next sample beaker to tower
¬ run lift with titration head to working position
¬ switch on stirrer
¬ start dosing device
¬ wait for 10 seconds
¬ start titrator
¬ wait for end of titration [EOD]
¬ switch off stirrer
¬ run lift with titration head to shift position
¬ wait for 3 seconds
¬ shift direction to decreasing numbers
¬ move special beaker to tower
¬ run lift to work position to immerse sensor
¬ switch on stirrer
¬ wait for 5 seconds
¬ switch off stirrer
¬ run lift to shift position
¬ wait for 3 seconds

¬ move special beaker to tower for conditioning
¬ run lift to work position to immerse sensor
------ settings of sampler functions --------------------




¬ on missing sample beaker, the next one is
selected automatically
------ stirring rates -----------------------------------------
¬ cannot be selected independently

------ settings for dosing devices ------------------------
------ action after manual stop -------------------------
¬ stop device 1

Advertising