25

9.

Lift the release handle on the

monochromator and engage the module
“catches” to the locking togs on the
monochromator. (Module not shown to
allow a good view of handle.)

Push monochromator and module
together firmly (with handle up) then
lower the release handle.

10.

When the catches are fully engaged to

the locking togs, push the release handle
down all the way.

This automatically engages the electrical
connector and fiber optic interface, and
maintains proper alignment of the
module to the instrument.

Carefully wind the fiber bundle clockwise
around the “saddle.” The 2-meter probe
takes one wrap. The 3-meter probe
(shown) takes 2 wraps.

The final assembly is as shown.

11.

Plug the AC power cord into a grounded AC outlet. A surge protector or Uninterruptible

Power Supply (UPS) is recommended for best operation.

12.

For “Direct Connection” the gray cable

from the instrument plugs directly to the
computer network jack. (Use the cable
supplied with the instrument.)

If using a network, use a non-crossover
type cable as listed in section 3.0. Plug
the RJ-45 network connector into a
functional network port.

If this requires approval from a network
administrator, it should be properly
approved for hookup.