46

▪▪▪▪▪▪▪

is done on an external standard, traceable to NIST.

The internal wavelength materials are used to maintain the wavelength registration of the
instrument. This is done by software. Final, tighter adjustments are later performed in Instrument
Calibration.

Select Wavelength Linearization from the
Diagnostics menu. The instrument will scan the
instrument reference.

Vision prompts to position the reference. Verify
that the Smart Probe is positioned as shown.

Verify the probe position and click “OK”.

Vision acquires spectra needed to perform
Wavelength Linearization.

No operator action is required at this point – this
screen is for information only.

Vision continues.

No operator action is required at this point – this
screen is for information only.