4 gain test, Gain test – Metrohm NIRS XDS SmartProbe Analyzer User Manual

Page 69

Advertising
background image

▪▪▪▪▪▪▪

67

In some cases, Photometric Test may fail and display the
message shown. This indicates that the photometric scale of
the instrument has shifted outside of Photometric Test
tolerances.

Failure may be due to several possible things:

The instrument Spectralon reference has become

soiled, and changed reflectivity, usually indicated by
the red spectrum being below the green spectral
limit,

The instrument needs to have Reference

Standardization re-run,

The probe window may be soiled,

Some combination of the above.

The user should verify proper test methods, cleanliness, and
Reference Standardization. If the Spectralon reference is
quite dirty, it may require replacement and subsequent re-
Reference Standardization.

7.2.4

Gain Test

The Gain Adjust feature can be a useful diagnostic tool, though it is not required for normal
operation. Technically, gain is never adjusted on the XDS Smart Probe Analyzer. The name of the test
comes from a capability required with older systems. With XDS, this program reports gain
information for the NIR and visible regions.

To start Gain Adjust, click on Diagnostics,
then Gain Adjust.

The instrument is connected and in
communication for this to function.