Instrument Diagnostics

Vision provides diagnostics for instrument setup, which must be performed before use of the
instrument for analysis. Following these diagnostics, another set of diagnostics is provided to
evaluate the ongoing performance of the instrument. These are explained in the sections that follow.

7.1

Setup Diagnostics

Before use of the XDS Analyzer, some steps must
be performed using Vision Software.

The “Options” for a given Project may be set. To
reach this menu, click on Configure, Options. The
menu shown at right is the default set of
selections. A brief explanation follows:

Instrument must stabilize before data acquisition:
This prevents spectral acquisition if the instrument
is cold.

Performance Test must pass before data
acquisition: This prevents the user from taking data
on a non-functional instrument.

Data Collection Method must match instrument
configuration:
Checked by default to prevent incorrect instrument
selection.

Run performance test after wavelength
linearization: Forces user to run test sequentially.
This is not necessary with XDS.

Use Auto-Linearization:
Maintains correct wavelength registration automatically, using internal wavelength materials to keep
instrument in precise adjustment over time.

Reference Standardization:
Used to create a virtual 100% reflectance reference, using a traceable photometric standard. This is
explained in the next section. Required for Method Transfer.

Blank Correction:
Not used on this model – applies only to XDS Rapid Liquid Analyzer.

Master Standardization:
This method is not used with XDS. Do not select.

Use Instrument Calibration (XDS only)
This is a method to adjust the instrument wavelength profile to an external, traceable wavelength
standard. It is checked as a default for XDS. Required for Method Transfer.

Use Window Correction:
This is used only on certain XDS Process Instruments. Do not select for XDS Smart Probe.