2 installation of the 757 va computrace stand, 1 mains cable and mains connection, 2 switching the instrument on/off – Metrohm 757 VA Computrace User Manual

Page 18

Advertising
background image

3 Installation

757 VA Computrace – Hardware

14

3.2

Installation of the 757 VA Computrace Stand

If the 757 VA Computrace Stand is connected to the power supply,

the instrument may not be opened or parts removed as there is a

danger of contact with live components. Before you open the 757 VA

Computrace Stand to change components or for maintenance or

repair work, always switch on the instrument by setting the mains

switch

12

to the ON position and then disconnect the mains cable

from the mains connection plug

13

of the 757 VA Computrace Stand !

3.2.1

Mains cable and mains connection

The instrument is supplied with one of three mains cables:

• 6.2122.020 with plug SEV 12 (Switzerland, …)
• 6.2122.040 with plug CEE(7), VII (Germany, …)
• 6.2133.070 with plug NEMA 5-15 (USA, …)

which are three-cored and fitted with a plug with an earthing pin. If a different plug

has to be fitted, the yellow/green lead (IEC standard) must be connected to protec-

tive earth (protection class 1).

Any break in the earthing inside or outside the instrument can make it

a hazard!

Plug the mains cable into mains connection plug 13 of the 757 VA Computrace

Stand (see Fig. 2).

3.2.2

Switching the instrument on/off

The 757 VA Computrace Stand is switched on and off using mains switch 12. When

the instrument is switched on, the pilot lamp 5 lights up.

Advertising