5 Technical data

757 VA Computrace – Hardware

58

Sweep rate

with voltage step 10 mV
CV: 0

… 30 V/s

SQW, DC: 0 … 20 V/s

DP, NP: 0 … 0.5 V/s

AC1, AC2: 0 … 0.02 V/s

Pulse amplitudes

AC1, AC2:

1 mV … 1 V

DP, NP:

–1 … 1 V

SQW: 0.15

mV

… 1 V

Current measurement

Current ranges

100 nA … 10 mA in 6 ranges

Current resolution

0.5 % of the current range

Minimum current I

min

5 pA

Maximum current I

max

35 mA

Integration times

0.1 … 20 ms

Multi-mode electrode MME (working electrode WE)

Designation

6.1246.020

Electrode types

DME (dropping mercury electrode)

HMDE (hanging mercury drop electrode)

SMDE (static mercury drop electrode)

Drop surface

0.15 ... 0.60 mm

2

(DME and SMDE)

Glass capillary

6.1226.030 (set of 10)

internal diameter = 0.05 mm

Mercury reservoir

6 mL ≅ 81.2 g; sufficient for ca. 200'000 Hg drops

Auxiliary power

inert gas (generally nitrogen N

2

); p = 1 ± 0.2 bar

Rotating disk electrode RDE (working electrode WE, option)

Construction

6.1246.000 Drive shaft + screw-on

6.1204.XXX Electrode tips

Electrode tips

6.1204.100 Ultra trace graphite

6.1204.110 glassy carbon (GC)

6.1204.120 platinum (Pt)

6.1204.130 silver (Ag)

6.1204.140 gold (Au) for Hg determination

6.1204.150 gold (Au) for As determination

Disk diameter

2.0 +0 / –0.05 mm

Radial eccentricity

≤ 0.2 mm

Regeneration

with 6.2802.000 Polishing kit

Rotational speed

200, 400, 600, ... , 3000 min

–1

Speed constancy

± 5 %

Reference electrode (RE)

Construction

double-junction; 6.0728.0X0 Ag/AgCl Ref. system +

6.1245.010 Electrolyte vessel to be filled by user

Reference system

Ag/AgCl/c(KCl) = 3 mol/L

Diaphragm

ceramic diaphragm; diameter = 3 mm