3 software settings – Metrohm 757 VA Computrace User Manual

Page 52

Advertising
background image

3 Installation

757 VA Computrace – Hardware

48

3.9.3 Software

settings

Before putting into operation the 757 VA Computrace Stand with the 813 Compact

Autosampler, the following settings have to be made in the "

757 VA Computrace

2.0" software program:

1 Set Dosimat parameters

• Click on

MAIN WINDOW

/ Settings / General settings

and select the

Hardware

tab.

• For each Dosimat connected to the remote interface of the 757 VA Com-

putrace Stand, check the

Dosimat no.

checkbox and enter the

Volume

of

the exchange unit installed on the Dosimat and the

Dose rate

of the Dosi-

mat.

• Define the addition or predose solutions for the desired method in the

DOSIMATS

window (procedure see Software Manual, section 5.2).

• Fill the exchange units of the Dosimats with the desired solutions and

make sure that there are no gas bubbles in the glass cylinders using the

button in the

DOSIMAT CONTROL

window (see section 6.2,

Software Manual).