5 reference electrode, 1 construction – Metrohm 757 VA Computrace User Manual

Page 40

Advertising
background image

3 Installation

757 VA Computrace – Hardware

36

3.5 Reference

electrode

3.5.1 Construction

The complete reference electrode (RE) 22 comprises two parts:

6.0728.0X0 Ag/AgCl reference system (101)

with ceramic diaphragm type D, diameter = 1 mm
6.0728.020

Reference system: Ag/AgCl/c(KCl) = 3 mol/L;

supplied in a holder filled with c(KCl) = 3 mol/L

as

standard

6.0728.010 Reference system: Ag/AgCl

supplied dry (option)

6.1245.010 Electrolyte vessel (102)

with ceramic diaphragm type D, diameter=3 mm; holds

a second electrolyte solution (bridge electrolyte) and

thus forms with the 6.0728.020 Reference system a

complete reference electrode in the so-called double

junction construction.

The construction of the reference electrode and the designations of the individual

parts are shown in Fig. 13.

22

102

109

110

111

101

103

104
105
106
107

108

Fig. 13: Construction of the reference electrode

22 Reference electrode

101 Reference system (6.0728.0X0)

102 Electrolyte vessel (6.1245.010)

103 Electrical connection for cable

"AE"

104 Vent opening

105 Ag/AgCl filling

106 Electrolyte compartment with

internal electrolyte

107 Diaphragm support made of

PCTFE

108 Diaphragm

109 Vent opening

110 Electrolyte compartment with

bridge electrolyte

111 Diaphragm

Advertising