4.1 Electrical safety

757 VA Computrace – Hardware

51

4 Safety


4.1

Electrical safety

While electrical safety in the handling of the 757 VA Computrace Stand is assured in

the context of the specifications IEC 61010-1 (protection class 1), the following

points should be noted:

• Mains connection

Setting

mains connection must be effected in accordance with the

instructions in section 3.2.1.

• Opening the instrument

When the 757 VA Computrace Stand is connected to the power

supply, the instrument may not be opened nor parts of them be

removed, otherwise there is a danger of coming into contact with

components which are live. Before you open the 757 VA Computrace

Stand to change components or for maintenance or repair work,

always switch on the instrument by setting the mains switch

12

to the

ON position and then disconnect the mains cable from the mains

connection plug

13

of the 757 VA Computrace Stand !

• Protection against static charges

Electronic components are sensitive to static charging and can be

destroyed by discharges. Before you touch any of the components

inside the 757 VA Computrace Stand, you should earth yourself and

any tools you are using by touching an earthed object (e.g. green

housing of the instrument or a radiator) to eliminate any static charges

which exist.