5 Technical data

757 VA Computrace – Hardware

59

Auxiliary electrode (AE)

Pt auxiliary electrode

6.0343.000 Platinum electrode

GC auxiliary electrode

(option)

6.1241.020 Electrode holder +

6.1247.000 Glassy carbon tip

Stirrer

Construction

6.1246.010 Drive shaft + screw-on

6.1204.090 Stirrer tip

Material

PTFE

Rotational speed

200, 400, 600, ... , 3000 min

–1

Speed constancy

± 5 %

Measuring vessels

6.1415.210

standard measuring vessel made of glass;

working volume = 10 ... 90 mL

6.1415.150

measuring vessel made of glass (option);

working volume = 5 ... 70 mL

6.1418.220

measuring vessel made of glass with thermostat

jacket (option); working volume = 12 ... 70 mL

6.1450.210

measuring vessel made of PFA (option)

working volume = 10 ... 90 mL

6.1456.210

measuring vessel made of glass for sample changer

operation (option); working volume = 10 ... 90 mL

6.1457.210

measuring vessel made of glass with thermostat

jacket for sample changer operation (option);

working volume = 10 ... 90 mL

Dummy Cell

Use

Checking the 757 VA Computrace Stand

Determination of the signal/noise ratio

Connections

AE

auxiliary electrode

RE

reference electrode

WE-L

working electrode linear mode

(RC element)

WE-D

working electrode differential mode

(peak/wave)

Inert gas (in general nitrogen N

2

)

Use

Operation of MME deaeration of sample solution

Required pressure

p = 1 ± 0.2 bar (this gas pressure results in a gas

flow rate of ca. 20 L/h)

Connection of Dosimats (Remote interface)

Type

665 or 765 Dosimat

Number

1...5

Plug

D-Sub with 25 pins

Manual operation

Dispensing, filling, adjustment of feed and filling rate