3.9 Connection of the 813 Compact Autosampler

757 VA Computrace – Hardware

43

3.9

Connection of the 813 Compact Autosampler

With the 813 Compact Autosampler connected to the 757 VA Computrace Stand,

max. 18 samples can be transferred to the measuring vessel at the 757 VA Com-

putrace Stand. After each measurement, the measuring vessel is rinsed by means

of two 772 Pump Units connected to a 731 Relay Box. For operation of this sample

changer and the automatic addition of standard addition and auxiliary solutions by

means of 765 Dosimats, the following instruments and accessory parts are needed

(see also section 6.2):

Quant.

Order no.

Instrument/Accessory

1

2.813.0020

813 Compact Autosampler for VA applications

1

2.731.0010

731 Relay Box

2

2.772.0010

772 Pump Unit

1 6.2141.150

Cable

813–757–731

1

6.5323.010

Rinsing equipment for VA

1…5 2.765.0010 765

Dosimat

1…5

6.3014.153

Exchange unit 5 mL (for addition solutions)

1…5

6.3014.213

Exchange unit 10 mL (for auxiliary solutions)

1 6.2141.080

Cable 757–2×765

or

1 6.9921.170

Cable 757–5×765

This section describes the procedure for the connection of 813 Compact Autosam-

pler, 731 Relay Box and 772 Pump Units. For the connection of 765 Dosimats, see

section 3.8.

3.9.1 Electrical

connection

Before any instruments are attached to the 757 VA Computrace Stand,

the 757 VA Computrace Stand must be

switched off using the mains

switch 12.

The 813 Compact Autosampler is connected to the socket "Remote" 15 of the 757

VA Computrace Stand with the optionally available

6.2141.150 Cable (see Fig. 15).

The second end of the 6.2141.150 cable is used to connect the 731 Relay Box. The

third end of the 6.2141.150 cable is used to connect 765 Dosimats with the

6.2141.080 cable (1...2 Dosimats) or the 6.9921.170 cable (1...5 Dosimats). At the

813 Compact Autosampler, "

Method 2" must be set (procedure see Instructions for

Use 813).
At the connection "DC1" of the 731 Relay Box, a 772 Pump Unit is connected as

si-

phoning pump, at the connection "DC2", a 772 Pump Unit is connected as rinsing

pump (see Instruction for Use 731 and 772). The output voltage of the 731 Relay

Box must be set to

+24 V (default setting, see section 2.3, Instruction for Use 713).

Additionally, the following settings for remote address selection must be done:

DC1: 9 (Siphoning pump)

DC2: A (Rinsing pump)