2 filling the mme with mercury – Metrohm 757 VA Computrace User Manual

Page 27

Advertising
background image

3.3 Multi-mode electrode (MME)

757 VA Computrace – Hardware

23

3.3.2

Filling the MME with mercury

When handling mercury, it is necessary to take special precautionary

measures. These are described in detail in section 4.2.

All actions involving the electrode and mercury vessels must be

performed in or over the drip pan 92 supplied (see Fig. 10).


The Hg reservoir 82 of the multi-mode electrode 17 is filled with mercury of the

highest degree of purity (mass fraction w ≥ 0.99999) as follows:

1 Prepare multi-mode electrode

• Unscrew locking ring 74 from slotted screw 75 (this gray PVC ring is

needed only to remove the slotted screws 75 or 80, see section 3.3.7 and

section 3.3.9).

• Turn slotted screw 75 in or out of the screw thread 77 using a suitable

coin until the contact surface of the black O-ring at the Plexiglas wall (thin,

black stripe) is just visible below the metal thread 77.

• Remove the plastic cap used as a transport safeguard from the retaining

nut 87.

• Undo retaining nut 87 fully and remove from screw thread 83.
• Place multi-mode electrode 17 with the capillary opening facing upwards

in the electrode holder 93 (see Fig. 9).

90

91

92

17

93

Hg

Fig. 9: Adding the mercury

17

Multi-mode electrode

(6.1246.0020)

92

Drip pan (6.2711.030)

90

Syringe (6.2816.020)

93

Electrode holder

(6.2615.030)

91

Needle (6.2816.030)