3 Installation

919 IC Autosampler plus

■■■■■■■■

13

This is how you install the needle and the capillary:

1 Removing the pressure screw

Loosen and remove the pressure screw (7-3) screwed onto the nee-
dle holder.

2 Inserting the needle

Insert the needle from above partway into the opening of the
needle holder (7-5).

Slide the PEEK ferrule (7-4) over the needle from above. The nar-
row side of the seal must face upwards.

3 Fastening the needle

Screw the pressure screw (7-3) into the needle holder. Lightly
push the needle upwards from below during the process.

Tighten the pressure screw in the needle holder by hand (do not
use tools!).

4 Connecting the capillary

Slide the 6.2744.070 PEEK pressure screw (7-2) over the end of
the capillary.

Manually tighten the PEEK pressure screw with the capillary to the
pressure screw of the needle holder. The capillary must be pushed
in while doing so.