4 operation and maintenance, 1 general, 2 filter – Metrohm 919 IC Autosampler plus User Manual

Page 33: General, Filter, Figure 17, Pump tubing connection – replacing the filter

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation and maintenance

919 IC Autosampler plus

■■■■■■■■

25

4 Operation and maintenance

4.1

General

Just like highly sensitive measuring instruments, Sample Processors require
proper care as well. Excess contamination of the instrument may result in
functional disruptions and a reduction in the lifetime of the otherwise
sturdy mechanics and electronics of the instrument.

Severe contamination can also have an influence on the measured results.
Regular cleaning of exposed parts can prevent this to a large extent.

Spilled chemicals and solvents must be removed immediately. In particular,
the connector strip (especially the power plug) should be protected from
contamination.

4.2

Filter

The 6.2821.130 filters (17-2) should be replaced every three months and
more frequently with higher backpressure.

1

2

3

Figure 17

Pump tubing connection – Replacing the filter

1

Tubing olive

2

Filter (6.2821.130)
Pack contains 10 pieces.

3

Filter screw

Replacing the filter

1 Unscrewing the filter screw

Unscrew the filter screw from the tubing olive.

2 Inserting the filter

Place the filter in the tubing olive and press it flat.

3 Installing the filter screw

Screw the filter screw back into the tubing olive.