5 filtration cell holder, 6 mounting the safety shield, Filtration cell holder – Metrohm 919 IC Autosampler plus User Manual

Page 25: Mounting the safety shield, Figure 10, Installing the filtration cell holder

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

919 IC Autosampler plus

■■■■■■■■

17

NOTE

Pump tubings are consumables. The lifetime of the pump tubings
depends on the contact pressure, among other factors.

3.5

Filtration cell holder

Figure 10

Installing the filtration cell holder

The filtration cell holder (6.2057.030) can be mounted on the side wall of
the tower, see above.

First remove the second and third screws from the bottom on the side
wall. Then fix the filtration cell holder in place with the two screws sup-
plied.

3.6

Mounting the safety shield

For reasons of safety, it is indispensable that you install the safety shield
(6.2751.110) supplied. A serious risk of injury exists if anyone reaches into
the working area of the instrument.

CAUTION

The 919 IC Autosampler plus may not be operated without a safety
shield.

Use the accompanying hex screws and the hex key to mount the safety
shield according to the following figure.

Advertising