1 connecting a dosing device, Connecting a dosing device – Metrohm 919 IC Autosampler plus User Manual

Page 29

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

919 IC Autosampler plus

■■■■■■■■

21

NOTE

When connecting MSB devices together, the following must be
observed:

Only one device of the same type can be used at a single MSB con-
nector at one time.

CAUTION

Exit the control software before you plug in MSB instruments. When it
is switched on, the control instrument automatically recognizes which
device is connected to which MSB connector. The operating unit or the
control software enters the connected MSB devices into the system
configuration (device manager).

MSB connections can be extended with the 6.2151.010 cable. The maxi-
mum connection length permitted is 15 m.

3.8.1

Connecting a dosing device

Three dosing devices can be connected to the instrument.

WARNING

If a Dosino is connected to the 919 IC Autosampler plus, then the con-
nection cable must be equipped with a T.2400.102 ferrite core. The fer-
rite core diminishes any interference voltages and thus ensures compli-
ance with the strict EMC standards in accordance with the applicable
technical standards, see Chapter "Technical data".

Proceed as follows:

1 Mounting the ferrite core

Fasten a T.2400.102 ferrite core to the Dosino connection cable near
to the plug.

2 Connecting a dosing device

Exit the control software.

Connect the connection cable of the dosing device to one of the
sockets marked with MSB on the rear of the control instrument.

Start the control software.

Advertising