6.1 Lift settings

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

919 IC Autosampler plus

6 Appendix

6.1

Lift settings

Adjusting the needle position

First move the lift to working height.

1 Open the safety shield

Loosen the lower two fastening screws of the safety shield using the
hexagon key provided, and tip up the safety shield.

2 Loosen the nut

Slightly loosen the nut below the lift head using a wrench.

3

2

Figure 18

Adjusting the needle

3 Adjust the needle holder

Turn the needle holder a few millimeters out of the lift head (turn
counterclockwise).

Now you can redefine the working height of the lift and precisely set
the needle position by turning the needle holder.

4 Tighten the nut

Using the wrench, fasten the nut again.

5 Close the safety shield

Tip down the safety shield and fix it with the fastening screws.