3.4 Installing the peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

919 IC Autosampler plus

3.4

Installing the peristaltic pump

1

2

3

4

5

6

7

8

1

9

10

3

6.2744.160

Figure 8

Installing the pump tubing

1

PEEK pressure screws, short
(6.2744.070)

2

Tubing olive (6.2744.030)

3

Stopper
The colors of the stoppers indicate the inner
diameter of the pump tubing.

4

Tubing cartridge (6.2755.000)

5

Contact pressure lever

6

Union nut

7

Adapter

8

Tubing olive

9

Pump tubing

10 Snap-action lever

Mount the pump tubing as follows:

1 Remove the tubing cartridge

Release the tubing cartridge from the cartridge holder by pressing
the snap-action lever and unhooking it from the mounting bolts
(5-1).

2 Insert the pump tubing

Press the contact pressure lever all the way down.

Place the pump tubing in the tubing cartridge. The stoppers (8-3)
must snap into the corresponding holder of the tubing cartridge.