2 connecting a stirrer or titration stand, Connecting a stirrer or titration stand, Figure 13 – Metrohm 919 IC Autosampler plus User Manual

Page 30: Connecting a dosing device, Figure 14, Connecting an msb stirrer, Figure 13 connecting a dosing device, You can use the following instruments, 1 connecting the stirrer or titration stand, Exit the control software

Advertising
background image

3.8 Connecting MSB devices

■■■■■■■■■■■■■■■■■■■■■■

22

■■■■■■■■

919 IC Autosampler plus

USB 1

Contr.

MSB 2

MSB 3

T.2400.102

Figure 13

Connecting a dosing device

3.8.2

Connecting a stirrer or titration stand

You can use the following instruments:

With built-in magnetic stirrer (stirring "from below"):

– 801 Stirrer
– 803 Ti Stand

Without built-in magnetic stirrer (stirring "from above"):

– 804 Ti Stand with propeller stirrer 802 Stirrer

Connect a stirrer or a titration stand as follows:

1 Connecting the stirrer or titration stand

Exit the control software.

Connect the connection cable of the magnetic stirrer or of the
titration stand to one of the sockets marked with MSB on the
rear of the control instrument.

804 Ti Stand only: Connect the propeller stirrer to the stirrer con-
nector (socket with stirrer symbol) of the titration stand.

Start the control software.

USB 1

Contr.

MSB 2

MSB 3

Figure 14

Connecting an MSB stirrer