4 error messages – Metrohm 795 KFT Titrino User Manual

Page 53

Advertising
background image

3.1 General rules

795 KFT Titrino

49

Status conditions of the global $S:

$S .Mode.KFT.QuickMeas:

Quick manual measurement from the initial status in mode KFT.

The instrument gives the status from which it has been stopped. The detailed

status information is therefore identical the information for the global status

$G.

Violation of monitored limits with action "end" give the status message

$S.Mode.KFT.Inac;EYYY.

3.1.4 Error messages

Error messages are added to the status messages and separated from them

by the sign ";".

E20

Check exchange unit.

Exit: Mount Exchange Unit (properly) or &m $S.

E21

Check electrode, short circuit.

Exit: Rectify fault or &m $S.

E22

Check electrode, break.

Exit: Rectify fault or &m $S.

E23

Division by zero.

Exit: The error message disappears on next startup or on recalculation.

E26

Manual stop.

Exit: The error message disappears on next startup.

E27

Stop V reached in KFT.

Exit: The error message disappears on next startup.

E28

Wrong object call up.

Exit: Send correct path for object. Start path at root.

E29

Wrong value or no value allowed.

Exit: Send correct value or call up new object.

E30

Wrong trigger, this trigger is not allowed or carrying-out of action not

possible.

Exit: Send correct trigger (exception: $D) or call up new object.

E31

Command is not possible in active status. Repeat command in inactive

status.

Exit: Send new command.

E32

Command is not possible during titration. Repeat command during the

conditioning phase or in inactive status.

Exit: Send new command.

E33

Value has been corrected automatically.

Exit: Send new command.

E34

Instrument at the end of the titration and sample data is edited; the instrument

at rest or editing during filling.

Exit: &m $S.