3.2 Remote control commands

795 KFT Titrino

67

3.2.2 Description of the remote control commands

3.2.2.1. Mode

$G, $S, $H, $C

Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and

continue with $C.

$G also serves to continue after inquiries of identifications and sample size af-

ter the start (see 3.2.2.26).

3.2.2.2. Mode.QuickMeas $G,

$S

Start and stop of a measurement in the basic mode with the parameters

(measured quantity, measuring input) of the current method. Corresponds to

the <meas/hold> key.

With an ongoing measurement, the current mode can be started. This stops

the measurement automatically.

3.2.2.3. Mode.Select KFT

Mode.KFTQuantity Ipol, Upol

Selection of the standard mode. Mode and the measured quantity belong to

the complete selection.

If a method is selected from the method memory, the nodes &Mode.Select

and &Mode.XXXQuantity are overwritten with mode and measured quantity of

the corresponding user method.

3.2.2.4. Mode.Name read

only

Name of the current method in the working memory. $Q sends 8 ASCII char-

acters. Standard methods carry the name ********. The node can be set

read + write, see 3.2.2.62.


3.2.2.5. Mode.Def.Formulas.1.Formula

EPX, CXX, RSX, +, –, ∗, /, (, )

Mode.Def.Formulas.1.TextRS

up to 8 ASCII characters

Mode.Def.Formulas.1.Decimal 0...2...5

Mode.Def.Formulas.1.Unit

up to 6 ASCII characters

Mode.Def.Formulas.1.Limits ON,

OFF

Mode.Def.Formulas.1.LoLim

0...±999 999

Mode.Def.Formulas.1.UpLim

0...±999 999

Mode.Def.Formulas.1.Output active,

pulse,

OFF

Mode.Def.Formulas.2.Formula

etc. up to .9

Entry of formulas. Rules for formula entry, see page 25.
Example: "(EP2–EP1)∗C01/C00"

In addition to the formula, a text for result output, the number of decimal

places and a unit for the result output can be selected. "No unit" is selected

with the blank string.

In place of "RSX", a result name may be entered (.TextRS). This name is out-

putted in the report full, short, scalc full and scalc srt. It is used for the result

and the corresponding mean value.

Advertising