Advertising
background image

3.2 Remote control commands

795 KFT Titrino

75

3.2.2.47. Config.Aux.DevName

up to 8 ASCII characters

Name of the instrument for connections with several units. It is advisable to

use only the letters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and

the numbers 0...9 (ASCII No. 48...57) when the function Setup.AutoInfo

(3.2.2.92) is used at the same time.

If a name has been entered, it will be printed out in the result report (full,

short).

3.2.2.48. Config.Aux.Prog read

only

Output of the program version.

The Titrino sends "795.0010" on requests with $Q.

3.2.2.49. Config.RSSet1 $G

Config.RSSet1.Baud

300, 600, 1200, 2400, 4800, 9600,

19200, 38400, 57600, 115200

Config.RSSet1.DataBit 7,

8

Config.RSSet1.StopBit

1, 2

Config.RSSet1.Parity even,

odd, none

Config.RSSet1.Handsh HWs, SWchar, SWline, none

$G sets all RS settings. The changes are performed only if the instrument is

inactive. After the setting of the interface parameters, wait at least 2 s to allow

the components to equilibrate.

Settings of the values for the data transmission via the RS interface: baud

rate, data bit, stop bit, parity and type of handshake, see also page 90ff. Baud

rates >9600 need a PC which is equipped accordingly (e.g. with 16550

component).

The setting of the values must be initiated with $G immediately after entry of

the values.

3.2.2.50. Config.ComVar.C30

with up to .C39, etc.

0... ±999999

Values of the common variables from C30 up to C39. Insert the common

variables directly or describe the determination results directly from the

method, see 3.2.2.7.

3.2.2.51. SmplData.Status ON, OFF

On/off switching of silo memory. When the silo memory is switched on, the

sample data are fetched from the lowest valid silo line.