Metrohm 795 KFT Titrino User Manual

Page 77

Advertising
background image

3.2 Remote control commands

795 KFT Titrino

73

3.2.2.32. UserMeth.List.1.Name read

only

UserMeth.List.1.Mode read

only

UserMeth.List.1.Quantity read

only

UserMeth.List.1.DosUnit read

only

UserMeth.List.1.Bytes read

only

UserMeth.List.1.Checksum read

only

for each method

List of the methods in the user method memory with the following character-

istics:

.Name:

Name of the method

.Mode: Mode

.Quantity: Measured

quantity

.DosUnit: Buret of the method

.Bytes:

Number of bytes of the user memory used by the method

.Checksum: Checksum of the method, see 3.2.2.61.

3.2.2.33. Config.Monitoring.Validation.Status ON,

OFF

Config.Monitoring.Validation.Interval 1...365...9999

Config.Monitoring.Validation.Counter

0...9999

Config.Monitoring.Validation.ClearCount $G

Monitoring of validation.

.Interval: Time interval in days for validation.

.Counter: Time counter in days since last validation.

.ClearCount: Clears the above counter.

3.2.2.34. Config.Monitoring.Service.Status ON,

OFF

Config.Monitoring.Service.Date YYYY-MM-DD

Monitoring of service interval.

3.2.2.35. Config.Monitoring.DiagRep ON,

OFF

Printing of system test report after each switching on of the Titrino.

3.2.2.36. Config.PeriphUnit.CharSet1

Epson, Seiko, Citizen, HP, IBM

Config.PeriphUnit.CharSet2

Selection of the character set and the graphics control characters for COM1

resp. COM2 of the Titrino.

IBM means the IBM character set following character set table 437 and IBM

graphics control characters. Select 'IBM' for work with the computer.

3.2.2.37. Config.PeriphUnit.RepToComport

1, 2, 1&2

Selection of COM of the Titrino where manually triggered reports should be

outputted.

3.2.2.38. Config.PeriphUnit.Balance Sartorius,Mettler,Mettler AT,

AND,Precisa

Selection of the balance type.

Advertising